- Flow rate at atm. pressure
- Up to 0.065 l/min
- Pressure head
- 60 mWg
- Suction height, self-priming
- 4.5 mWg
- Nominal stroke volume
- 50 – 200 μl
What is KNF “Versatile Dosing,” and how does it simplify mechanical design?
Featured in KNF’s next-generation Intelligent Pumps, “Versatile Dosing” replaces manual mechanical adjustments with electronic settings, reducing adjustment errors and simplifying pump integration.
- Electronic dosing control: Set dosing behavior through software without mechanical adjustments.
- On-the-fly adjustment: Change dosing parameters from shot to shot.
- Cavitation prevention: A stepper motor and actively controlled valves allow separate intake and discharge speeds to accommodate liquid viscosity and help prevent cavitation.
Which communication interfaces does the KNF FD 1.200 support, and how easy is it to integrate?
The FD 1.200 offers several control and installation options:
- Serial communication: A 5-pin interface uses UART TTL as standard, with RS-232 and CAN bus available as project options. I²C availability requires project confirmation.
- I/O control: A separate 12-pin interface handles digital and analog inputs and outputs independently of serial communication.
- Faster development: A USB connection and PC GUI support parameter testing. Validated settings can be sent to KNF for a customized pump configuration.
- Mechanical mounting: Three mounting fixture options provide installation flexibility.
What are the key performance and operating specifications of the FD 1.200?
The stepper motor supports partial strokes and adjustable stroke speed for precise dosing. A stiff diaphragm with rigid support helps maintain repeatability as pressure changes.
| Parameter | Specification | Notes |
|---|---|---|
| Shot volume | 50–200 µl | For small-volume dosing |
| Maximum flow rate | 40 ml/min in dosing mode 65 ml/min in priming mode |
Maximum shot frequency: 3.33 shots/s |
| Setpoint accuracy / repeatability (CV) | ±2% / ≤0.5% | Factory calibration: water at 25 °C and 0 bar gauge |
| Maximum pressure head | Outlet: 6 bar (85 psig) Inlet: 2 bar (28.5 psig) |
|
| Dry suction height | ≥4.5 mWg | At a 200 µl shot volume |
| Electrical supply | 24 VDC (±10%), 36 W | Maximum load current: 1.5 A |
| IP protection | IP20 as standard | IP54 available on request |
Which wetted materials are available for corrosive media in the FD 1.200?
The FD 1.200 has a PP housing and a chemically resistant PPS pump head. Two standard material variants cover different liquid compatibility needs:
| Variant | Application considerations | Wetted elastomers and diaphragm |
|---|---|---|
| RP | For liquids compatible with EPDM | O-rings, diaphragm and valves: EPDM |
| RT | For applications requiring PTFE and FFKM chemical resistance; compatibility also depends on the FKM O-rings | O-rings: FKM Diaphragm: PTFE Valves: FFKM |
What affects dosing accuracy in the FD 1.200?
The ±2% setpoint accuracy applies under factory calibration conditions. Suction height, outlet pressure, viscosity, temperature and operating time can affect the delivered volume:
| Influence on accuracy | RP | RT |
|---|---|---|
| Suction height | 0 / −2% per mWg | 0 / −2% per mWg |
| Outlet pressure at 200 µl | 0 / −0.2% per mWg | 0 / −0.2% per mWg |
| Outlet pressure at 50 µl (minimum shot volume) | 0 / −0.6% per mWg | 0 / −0.6% per mWg |
| Liquid viscosity: 0.3–150 mPa·s | 0 / −3% | 0 / −5% |
| Liquid temperature: 5–80 °C | ±2% | ±2% |
| Ambient and liquid temperature: 5–50 °C | ±3% | ±3% |
| Drift at 200 µl after 54 million cycles | ±4% | ±4% |
| Drift at 50 µl after 54 million cycles* | ±8% | ±8% |
*Primarily relevant to small target volumes of 50 µl.
What are the benefits of the actively actuated valves in the FD 1.200?
The actively actuated inlet and outlet valves in the FD 1.200 provide precise valve control for accurate dosing and flexible pump operation. Key benefits include:
- Increased dosing precision
- Bi-directional flow
- Backflow-tight operation in both forward and reverse directions
What applications is the FD 1.200 suitable for?
The FD 1.200 is suited to medical diagnostics, laboratory automation and precision liquid dosing. Depending on liquid compatibility, applications also include food ingredient dosing, cleaning and disinfection, and industrial ink delivery.
Electronic dosing settings, adjustable intake and discharge speeds, and multiple control interfaces help engineers adapt the pump to their equipment.
FD 1.200 Features & Specifications
Technical Details
| Flow rate | 0.065 l/min at atm. pressure |
|---|---|
| Weight | 0.7 kg |
| Shot volume range | 50 – 200 μl |
| Max. shot frequency (shots per second) | 3.33 |
| Max. flow rate: dosing mode | 40 ml/min |
| Max. flow rate: priming mode | 65 ml/min |
| Set point accuracy | ± 2% RD |
| Repeatability (CV) | ≤ 0.5% |
| Calibration | Calibration in factory @ 0 bar g / 25 °C with water |
| Suction height, dry | ≥ 4.5 mWg @ 200 μl shot volume |
| Max. outlet pressure head | 60 mWg |
| Max. inlet pressure head | 20 mWg |
| Permissible media temperature | +5° to +80°C |
| Max. media temp. for hot sterilization | 95°C for max. 40 minutes |
| Permissible ambient temperature | +5° to +50°C |
| IP protection factor | IP20 (IP54 on request) |
| Operating voltage | 24 ± 10% VDC |
| Power | 36 W |
| I load max | 1.5 A |
Accessories & Spare Parts
| Title | Description | Order No. |
|---|---|---|
| Starter-Kit cable USB – UART TTL | USB Connecting cable to parametrize the FD 1.200 System with the KNF Connect Software, open wire ends for power | 358830 |
| Starter-Kit cable USB – RS-232 | USB Connecting cable to parametrize the FD 1.200 System with the KNF Connect Software, open wire ends for power | 365095 |
| Filter: FS 60 T PVDF / FS 60 X PEEK | KNF filters protect pumps and upstream equipment from particles, crystals and fibres; mesh opening: PVDF 70 μm, PEEK 35 μm | PVDF: 346932 / PEEK: 323625 |