Micro Diaphragm Gas Pumps
NVC 830.3
- Flow rate at atm. pressure
- Up to 2.6 l/min
- Vacuum
- 55 mbar abs.
- Pressure
- 0 bar g
What are the benefits of “Closed-Loop Vacuum Control” in the KNF NVC 830.3?
KNF developed “Closed-Loop Vacuum Control” to continuously monitor vacuum levels through integrated sensing and feedback and adjust pump operation accordingly. In the NVC 830.3, the system regulates pressure toward a defined setpoint, offering the following benefits for system integration:
- Pressure regulation: continuous monitoring allows the pump to correct deviations from the setpoint. The datasheet specifies sensor accuracy of ±15 mbar and drift of 1% per year; these are sensor specifications, not a guarantee of overall system control accuracy.
- Dynamic response: the controller adjusts pump operation when system pressure changes, helping stabilize the process within the pump’s operating limits.
- Demand-based operation: output adjusts to the vacuum demand, which can reduce unnecessary full-speed running and associated power consumption and noise.
- Simpler system design: integrated vacuum regulation can reduce the need for a separate controller or pressure switch dedicated to that function.
What are the key features and advantages of the NVC 830.3?
The NVC 830.3 combines a compact dual-head diaphragm pump with integrated vacuum control:
| Feature | Function and benefits |
|---|---|
| Closed-loop vacuum control | Monitors vacuum and adjusts pump output to maintain the selected setpoint as system demand changes. |
| Digitally customizable motor | BLDC control parameters and interface settings can be configured to match the customer’s OEM system requirements. |
| Compact performance | Provides vacuum generation in a compact package, supporting integration into equipment with limited installation space. |
| Clean gas transfer | Oil-free operation avoids lubricant contamination of the pumped gas. |
| Material and operating options |
|
What are the main performance and motor specifications of the NVC 830.3?
| Specification | 6 V version | 12 V version | 24 V version |
|---|---|---|---|
| Voltage (V) | 6 | 12 | 24 |
| Flow rate at atmospheric pressure (l/min) | 1.2 | 2.4 | 2.6 |
| Ultimate vacuum (mbar abs.) | 55 | 55 | 55 |
| Motor | Integrated brushless DC, 2-wire / 4-wire / 5-wire configurations, 6 – 24 V | ||
| Imax (A) | 0.70 | 0.75 | 0.55 |
| Communication interface | UART TTL, 3.3 V level (UART RS232, I2C or CAN Bus on request) | ||
How many operating modes does the NVC 830.3 offer?
- Mode 1: Fixed Set Point
- Operation: the target vacuum is configured before delivery.
- Application: suits equipment requiring one vacuum setpoint; the pump evacuates to that level and maintains it through closed-loop control.
- Mode 2: Controllable Set Point
- Operation: an analog input changes the vacuum setpoint. The datasheet’s 0 – 5 V mapping is an example: 1 V corresponds to -200 mbar gauge and 5 V to -1,000 mbar gauge.
- Feedback: a PWM output reports the measured vacuum level to the host controller.
The mapping illustrates the control signal settings; actual vacuum performance depends on the pump specifications and system conditions. The NVC 830.3 has a specified ultimate vacuum of 55 mbar absolute, which should be considered when selecting the target vacuum level.
- Mode 3: Communicate
- Operation: connects to an external controller through UART TTL serial communication.
- Integration: supports real-time start/stop commands, multiple setpoints and PID parameter changes. Command libraries in several programming languages support integration into digitally controlled laboratory and semiconductor equipment.
What applications is the NVC 830.3 suitable for?
Oil-free gas transfer and integrated vacuum regulation support applications including:
- Medical and laboratory equipment: analytical instruments, in vitro diagnostics and laboratory vacuum filtration.
- Industrial equipment: semiconductor processes, inkjet printing and vacuum technology.
- Food and beverage production equipment, with materials selected for the application.
- Other applications: automotive systems, security and defense equipment, gas analysis and fuel cells.
NVC 830.3 Features & Specifications
Technical Details
| Flow rate | 2.6 l/min at atm. pressure |
|---|---|
| Ultimate vacuum | 55 mbar abs. |
| Permissible media temperature | +5° to +40°C, extended temperature on request |
| Permissible ambient temperature | +5° to +40°C, extended temperature on request |
| Weight | 0.258 kg |
| Sensor accuracy / drift | ± 15 mbar / 1% per year |
| Communication interface | UART TTL 3.3V Level (UART RS232, I2C or CAN Bus on request) |
Accessories & Spare Parts
| Title | Description | Order No. |
|---|---|---|
| Spare Part Kit | One kit per head containing valve and diaphragm. For a double-headed pump, two kits are required | On request |
| Eccentric | On request | |
| Silencer | 338498 | |
| Rubber-Bonded Metal | On request | |
| Serial Interface Adapter | 347669 | |
| Filter | 316422 |